4-24 | Model 5100 HD Analyzer
Data Type
Description
MB Starting
Regs
MB Reg Size
float
Relay 4 Value. The value used to evaluate the
relay function.
167
2
boolean
Master enable for all relays. 0=Disabled,
1=Enabled
169
1
char[20]
Analyte 1 Name
170
10
float
Analyte 1 Concentration
180
2
float
Analyte 1 band high limite band 1
182
2
float
Analyte 1 band low limite band 1
184
2
float
Analyte 1 band high limite band 2
185
2
float
Analyte 1 band low limite band 2
188
2
float
Analyte 1 band high limite band 3
190
2
float
Analyte 1 band low limite band 3
192
2
float
Analyte 1 band high limite band 4
194
2
float
Analyte 1 band low limite band 4
196
2
char[20]
Analyte 2 Name
198
10
float
Analyte 2 Concentration
208
2
float
TEC Temperature Value
210
2
int16
Heater Actuator Type. 0=Reverse, 1=Forward.
212
1
float
Heater Deviation Tolerance. Ignored if 0.
Otherwise high or low limit alarm are set when
value is out bounds of 1 deviation
213
2
float
Heater Deriviative Time
215
2
float
Heater Duty Cycle
217
2
float
Heater Integral Time
219
2
float
Heater Proportional Band
221
2
float
Heater Set Point
223
2
float
TEC Sample Time
225
2
float
TEC hysterisi
227
2
float
Electronics Temperature
229
2
float
Cell RTD1 Filtered value
231
2
float
Cell Set Point
233
2
float
Cell Deviation Tolerance. Ignored if 0. Oth-
erwise high or low limit alarm are set when
value is out bounds of 1 deviation
235
2
float
Calibration temperature for concentration
compensation.
237
2
float
Calibration pressure for concentration com-
pensation
239
2
float
Max TEC temperature
241
2
float
Min TEC temperature
243
2
uint
Interval in minutes to save the log data
245
2
char[10]
Concentration Unit of Measure for analyte 2
247
5
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...