User Interface
|
4-31
Data Type
Description
MB Starting
Regs
MB Reg Size
float
Power correction alpha
634
2
int16
Beta value for wir filter
636
1
float
Power correction Z value
637
2
float
Duty cycle for the sample heater
639
2
float
Power correction Z value 2
641
2
float
Laser power of sample channel
643
2
float
Laser power reference of sample channel
645
2
float
Laser power of sample channel 2
647
2
float
Laser power reference of sample channel 2
649
2
float
concentration span
651
2
float
concentration offset
653
2
float
concentration span 2
655
2
float
concentration offset 2
657
2
float
span ratio
659
2
float
span ratio 2
661
2
int16
status of span check
663
1
int16
status of span check
664
1
float
zero check
665
2
float
zero check
667
2
int16
status of zero check
669
1
int16
status of zero check
670
1
float
For span check - warning limit
671
2
float
For span check - alarm limit
673
2
float
Zero tolerance (of calib range)
675
2
int16
(NOT USED)
677
1
int16
(NOT USED)
678
1
int16
(NOT USED)
679
1
int16
(NOT USED)
680
1
float
(NOT USED)
681
2
float
(NOT USED)
683
2
float
(NOT USED)
685
2
int16
Auto calibration. 0=disabled, 1=enabled.
687
1
int16
0=Zero+Span, 1=Zero, 2=Span
688
1
int16
The last day that the calibration was triggered
by the scheduler
689
1
int16
Cal period in second
690
1
boolean
1=perform autocal immediately
691
2
float
(NOT USED)
693
2
float
(NOT USED)
695
2
int16
0=do not adjust auto cal; 1=adjust auto cal
697
1
int16
(NOT USED)
698
1
int16
System State (Laser 2). 100=Not Ready,
201=Measure Run, 502=Idle, 600=diag...
699
1
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...