4-29
Data Type
Description
MB Starting
Regs
MB Reg Size
float
Heater Proportional Band
492
2
float
Cell Sample Time
494
2
float
Heater Deriviative Time
496
2
float
Heater Duty Cycle
498
2
float
Heater Integral Time
500
2
float
Heater Proportional Band
502
2
float
Heater Set Point
504
2
float
Ref Sample Time
506
2
int16
Midpoint of the DAC range for TEC control
508
1
float
Pressure compensation coefficient 1
509
2
float
Pressure compensation coefficient 2
511
2
float
Pressure compensation coefficient 3
513
2
float
Pressure compensation coefficient 4
515
2
float
Analyte 1 Concentration in other unit
517
2
char[20]
2nd Concentration Unit of Measure for analyte
1
519
10
uint
Data Valid. 0=Not Valid, 1=Valid
529
2
int16
0=Line-lock mode, 1=Sample, 2=Calib
531
1
int16
Pressure and temp compensation. 0=none,
1=pressure, 2=temperature, 3=both
532
1
float
Process seal pressure span
533
2
float
Process seal pressure offset
535
2
float
Process seal pressure IIR
537
2
float
Process seal max
539
2
float
Process seal min
541
2
float
Sample channel's threshold
543
2
float
Sample channel 2's threshold
545
2
float
0=single, 1=dual concentration from one
spectrum
547
2
float
TEC Temperature Value
549
2
int16
Heater Actuator Type. 0=Reverse, 1=Forward.
551
1
float
Heater Deviation Tolerance. Ignored if 0.
Otherwise high or low limit alarm are set when
value is out bounds of 1 deviation
552
2
float
Heater Deriviative Time
554
2
float
Heater Duty Cycle
556
2
float
Heater Integral Time
558
2
float
Heater Proportional Band
560
2
float
Heater Set Point
562
2
float
TEC Sample Time
564
2
float
TEC hysterisi
566
2
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...