Maintenance
|
6-7
30
141
r
float
ANAOUT2
135
2
Current Output 2 -
142
rw
float
ANAOUT2O
137
2
Current Output 2 Offset -
143
rw
uint16
ANAOUT2P
139
1
Current Output 2 Parameter-
144
rw
float
ANAOUT2S
140
2
Current Output 2 Span -
145
rw
int16
ANAOUT2T
142
1
Current Output 2 Output Type. 0=4-
20 mode, 1=0
-
20 mode.
If overange enable (ANAOUT2OR =
1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or
low values can also be set by fault
value, See Fault option (ANAOUT2F)
-
146
rw
uint16
ANAOUT2OR
143
1
Current Output 2
Over range Enable.
0=disabled, 1=enabled
147
rw
uint16
ANAOUT2F
144
1
Current Output 2 Fault Option. 0=No
fault, 1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT2T = 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
148
r
uint16
RELAY1
145
1
Relay 1. 0=OPEN, 1=CLOSED
149
rw
uint16
RELAY1D
146
1
Relay 1 Delay. Holds the relay output
for this set amount of time.
150
rw
uint16
RELAY1F
147
1
Relay 1 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
151
rw
uint16
RELAY1S
148
1
Relay 1 Source. The source
parameter number used to evaluated
the relay function.
152
rw
float
RELAY1V
149
2
Relay 1 Value. The value used
to
evaluate the relay function.
153
r
uint16
RELAY2
151
1
Relay 2. 0=OPEN, 1=CLOSED
154
rw
uint16
RELAY2D
152
1
Relay 2 Delay. Holds the relay output
for this set amount of time.
155
rw
uint16
RELAY2F
153
1
Relay 2 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
156
rw
uint16
RELAY2S
154
1
Relay 2 Source. The source
parameter number used to evaluated
the relay function.
157
rw
float
RELAY2V
155
2
Relay 2 Value. The value used to
evaluate the relay function.
158
r
uint16
RELAY3
157
1
Relay 3. 0=OPEN, 1=CLOSED
159
rw
uint16
RELAY3D
158
1
Relay 3 Delay. Holds the relay output
for this set amount of time.
160
rw
uint16
RELAY3F
159
1
Relay 3 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
161
rw
uint16
RELAY3S
160
1
Relay 3 Source. The source
parameter number used to evaluated
the relay function.
162
rw
float
RELAY3V
161
2
Relay 3 Value. The value used to
evaluate the relay function.
163
r
uint16
RELAY4
163
1
Relay 4. 0=OPEN, 1=CLOSED
31
164
rw
uint16
RELAY4D
164
1
Relay 4 Delay. Holds the relay output
for this set amount of time.
165
rw
uint16
RELAY4F
165
1
Relay 4 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
166
rw
uint16
RELAY4S
166
1
Relay 4 Source. The source
parameter number used to evaluated
the relay function.
167
rw
float
RELAY4V
167
2
Relay 4 Value. The value used to
evaluate the relay function.
168
rwf
boolean
RELAYENA
169
1 Master enable for all relays.
0=Disabled, 1=Enabled
169
rwf
char[20]
ALYTE1
170
10
Analyte 1 Name
170
r
float
ALYTE1CON
180
2
Analyte 1 Concentration
171
rw
float
CONCLIMITH1
182
2
Analyte 1 band high limite band 1
172
rw
float
CONCLIMITL1
184
2
Analyte 1 band low limite band 1
173
rw
float
CONCLIMITH2
186
2
Analyte 1
band high limite band 2
174
rw
float
CONCLIMITL2
188
2
Analyte 1 band low limite band 2
175
rw
float
CONCLIMITH3
190
2
Analyte 1 band high limite band 3
176
rw
float
CONCLIMITL3
192
2
Analyte 1 band low limite band 3
177
rw
float
CONCLIMITH4
194
2
Analyte 1 band high limite band 4
178
rw
float
CONCLIMITL4
196
2
Analyte 1 band low limite band 4
179
rwf
char[20]
ALYTE2
198
10
Analyte 2 Name
180
r
float
ALYTE2CON
208
2
Analyte 2 Concentration
181
r
float
TECTEMP
210
2
TEC Temperature Value
182
rwf
int16
TECAT
212
1
Heater Actuator Type. 0=Reverse,
1=Forward.
183
rwf
float
TECDEVT
213
2 Heater Deviation Tolerance. Ignored if
0. Otherwise high or low limit alarm
are set when value is out bounds of 1
deviation
184
rwf
float
TECDT
215
2 Heater Deriviative Time
185
r
float
TECHDC
217
2 Heater Duty Cycle
186
rwf
float
TECIT
219
2 Heater Integral Time
187
rwf
float
TECPB
221
2 Heater Proportional Band
188
rwf
float
TECSP
223
2 Heater Set Point
189
rwf
float
TECST
225
2
TEC Sample Time
190
rwf
float
TECHYS
227
2
TEC hysterisi
191
r
float
ELECTEMP
229
2
Electronics Temperature
192
r
float
CELLTEMP
231
2
Cell RTD1 Filtered value
193
rwf
float
CELLSP
233
2 Cell Set Point
194
rwf
float
CELLDEVT
235
2 Cell Deviation Tolerance. Ignored if 0.
Otherwise high or low limit alarm are
set when value is out bounds of 1
deviation
195
rwf
float
NORMTEMP
237
2 Calibration temperature for
concentration compensation.
196
rwf
float
NORMPRES
239
2 Calibration pressure for concentration
compensation
197
rwf
float
MAXTEMP
241
2
Max TEC temperature
198
rwf
float
MINTEMP
243
2
Min TEC temperature
199
rw
uint
LOGSAVEINT
245
2 Interval in minutes to save the log
data
200
rw
char[10]
CONCUNIT2
247
5 Concentration Unit of Measure for
analyte 2
201
rw
uint
ALYTE1RES
252
2 Decimal places to use for displaying
analyte 1
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...