2-4 | Model 5100 HD Analyzer
Sample Temperature
-20C to +150C (-4 F to +302 F) Gas-phase samples
only
Ambient temperature
-20
o
C to +50
o
C (-4
o
F to +122
o
F)
Storage temperature
-25 C to +80 C (-13 F to +176 F)
Ambient humidity
0-90%, non-condensing
Physical
880 mm x 674 mm x 302 mm (34.6” x 26.5” x 11.9”),
60 kgs (132 lbs)
Enclosure material
300 series stainless steel
Maximum altitude
2000 meters
Pollution Degree
2
Ingress protection (IP)
IP65 and NEMA Type 4X
Sample flow rate
1-10 lpm (2-20 SCFH)
Sample gas
Natural gas, Hydrogen, Carbon Dioxide, Propane,
Oxygen, and others pending application review
Sample wetted materials
Stainless steel, SiO2 glass, Gold, Teflon, EPDM
rubber
Electrical Classifications
This analyzer will be certified with regard to
electrical safety for Pollution Degree 2 (normally
non-conductive environmental pollution occurs
with occasional condensation) and Installation
Category II (local level transients, less than
those found at power distribution levels). Mains
supply voltage fluctuations are not to exceed 10
percent of the nominal supply voltage. Other ap-
provals will include acceptance for use in
NEC/CEC Class I Division 2 Groups A, B, C, D T3
and world-wide group II Zone 2 IIC T3 classified
hazardous areas.
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...