GASGUARD
®
TE10 Distribution System Manual
Air Products and Chemicals, Inc.
Page 4 - 4
Figure 4.3:
Power Supply Terminal Connector
The power requirements are as follows:
100-240 VAC @ 150 VA maximum.
Overvoltage protection:
Recommended; may be required by local code
Sizing:
25% (minimum) over required load (add all cabinet
loads
and
divide
by
0.75)
NOTE: Power wiring must be sized to deliver the required voltage at
the rated current. Voltages should be checked at each cabinet after
installation to ensure proper levels.
C irc u it
B rea k e r
A C S urg e
P ro te cto r
H o t
N eu tra l
G ro u n d
Содержание GASGUARD TE10
Страница 2: ......
Страница 12: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page i x...
Страница 70: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page 3 4...
Страница 87: ...Section 4 Electrical Connections Revision August 2005 Page 4 7 Figure 4 5 Conduit and Conduit Seals...
Страница 92: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page 4 12...
Страница 94: ......
Страница 154: ......
Страница 166: ......
Страница 174: ......
Страница 180: ......
Страница 188: ......
Страница 189: ...Appendix B Revision November 2004 Page B 1 Appendix B This Appendix contains the Nitrogen N2 Material Safety Data Sheet...
Страница 190: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page B 2...
Страница 191: ...Appendix B Revision November 2004 Page B 3...
Страница 192: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page B 4...
Страница 193: ...Appendix B Revision November 2004 Page B 5...
Страница 194: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page B 6...