62
Installation and Verification Manual
3
Performance Tests
Recall Default Settings
The Performance Tests require the OmniBER OTN to be set to a
pre-defined (default) state at the beginning of each test.
1
Press
<Menu>
, choose
Tx/Rx > Stored Settings
then press
<Select>
.
2
Select
Default
then
Recall.
3
Select
OK
in the Warning box and then wait a few seconds for
instrument to reconfigure to the default settings. The recall is
complete when the Warning box diappears.
4
Select
Close
in the
Tx/Rx - Stored Settings
window.
To recall default settings