Performance Tests
3
Installation and Verification Manual
121
2
Press
<Menu>
, select
Tx/Rx > Transmitter
and set the
Signal Mode
to
PDH/DSn
. Now set the
Signal Rate
to
2Mb/s, Termination 75 ohms (Unbal)
.
3
Connect the
2 Mb/s BAL OUT
port to the Oscilloscope CHAN 1,
using the E2622A Termination Adapter Probe.
.
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
2 Mb/s UnBalanced Output
BNC to BNC cable
E2622A
75/50
Ω
2 Mb/s UnBal OUT
panther3_iv.book Page 121 Wednesday, January 15, 2003 12:03 PM