138
Installation and Verification Manual
3
Performance Tests
4
Load the
STS3_ANSI.msk
test mask and check the eye meets the
mask (
see example).
5
Press
<Menu>
, select
Tx/Rx > Transmitter settings
and set
the
Signal Mode
to
SDH
. Now set the
Signal Rate
to
STM-1,
and
Interface
to
Electrical
.
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
STS-3/STM-1e Output
BNC to BNC cable
E2622A
75/50
Ω
STS-3/STM-1e Output
STS-3 Eye Mask Test
panther3_iv.book Page 138 Wednesday, January 15, 2003 12:03 PM