Installation and Verification Manual
2
Installation (Getting Started)
Instrument Reboot (Cold Start)
An instrument “cold start” routine is provided to reset the
instrument in the event of an unplanned hardware or firmware
event. A cold start reboots the instrument and restarts the
instrument using a default configuration file. Performing a cold
start erases existing configuration information. Current menu
settings will be lost and the instrument will power up in its
default state.
1
Switch the instrument off and wait a few seconds.
2
Switch the instrument on and as the instrument boots up,
look carefully at the display.
3
Wait for the “Starting instrument” text to be displayed. After
a few seconds start to repeatedly press the
<Menu>
key until
the Agilent splash screen appears with an options menu in
the top left corner of the display.
The following options are available:
3
Cold start.
5
Normal start.
4
Press 3 on the numeric keypad to select cold start. The unit
will then continue with the boot up process.
5
When the boot-up procedure is complete, the instrument
displays a dialog box with the message: “Instrument reset to
default settings.”
To perform a cold start
panther3_iv.book Page 56 Wednesday, January 15, 2003 12:03 PM
Содержание OmniBER J7230A
Страница 2: ...sdh_Lynx2 book Page 148 Wednesday April 17 2002 12 49 PM...
Страница 10: ...8 Installation and Verification Manual panther3_iv book Page 8 Wednesday January 15 2003 12 03 PM...
Страница 172: ...170 Installation and Verification Manual panther3_iv book Page 170 Wednesday January 15 2003 12 03 PM...
Страница 173: ...sdh_Lynx2 book Page 148 Wednesday April 17 2002 12 49 PM...