Performance Tests
3
Installation and Verification Manual
129
3
Connect the
DS-3 OUT
port to the Oscilloscope CHAN 1,
using the E2622A Termination Adapter Probe.
4
Load the
DS3_ANSI.msk
test mask and check the Positive
and Negative pulses are within the mask (see examples).
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
DS3 Output
BNC to BNC cable
E2622A
75/50
Ω
DS3 O/P
N O T E
To test Negative pulse set the Scope
Mask Test Set-up
panel to
Invert
Mask
panther3_iv.book Page 129 Wednesday, January 15, 2003 12:03 PM
Содержание OmniBER J7230A
Страница 2: ...sdh_Lynx2 book Page 148 Wednesday April 17 2002 12 49 PM...
Страница 10: ...8 Installation and Verification Manual panther3_iv book Page 8 Wednesday January 15 2003 12 03 PM...
Страница 172: ...170 Installation and Verification Manual panther3_iv book Page 170 Wednesday January 15 2003 12 03 PM...
Страница 173: ...sdh_Lynx2 book Page 148 Wednesday April 17 2002 12 49 PM...