Performance Tests
3
Installation and Verification Manual
119
.
4
Load the
2Mb_ITU_120.msk
test mask and check the Positive
and Negative pulses are within the mask (see examples).
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
2 Mb/s Balanced Output
3-pin Siemens cable
E2621A
2 Mb/s Bal OUT
Set
to
120
Ω
N O T E
T
o test Negative pulse set the Scope
Mask Test Set-up
panel to
Invert
Mask.