Performance Tests
3
Installation and Verification Manual
137
1
Recall the OmniBER OTN’s default setting as shown on
2
Press
<Menu>,
select
Tx/Rx > Transmitter Settings
and set
the
Signal Mode
to
SONET
. Now set the
Signal Rate
to
STS-3
.
3
Connect the
155 Mb/s OUT
port to the Oscilloscope CHAN 1, using
the E2622A Termination Adapter Probe
.
To test the 155 Mb/s output pulse mask
panther3_iv.book Page 137 Wednesday, January 15, 2003 12:03 PM