![Agilent Technologies G7129A Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/agilent-technologies/g7129a/g7129a_user-manual_2867753042.webp)
Agilent InfinityLab LC Series Vialsamplers User Manual
42
2
Site Requirements and Specifications
Performance Specifications
Performance Specifications (G7129B)
Table 3
Performance Specifications G7129B
Type
Specification
Comments
Method/Conditions
Injection range
0.1 – 20 µL in 0.1 µL increments
(default)
0.1 – 40 µL in 0.1 µL increments
(if 40 µL loop is installed)
0.1 – 120 µL in 0.1 µL incre-
ments with 1290 Infinity large
volume injection kit (hardware
modification required)
0.1 – 100 µL in 0.1 µL (if 100 µL
loop and 100 µL head is
installed)
Up to 1500 µL (with 1400 µL
multi-draw kit and 100 µL ana-
lytical head)
up to 130 MPa (1300 bar,
18854 psi)
up to 60 MPa (600 bar,
8702 psi)
Injection precision
<
0.25 % RSD of peak areas from
5 µL to 100 µL
Measured caffeine
Pressure range
Up to 130 MPa (1300 bar,
18854 psi)
Sample viscosity
range
0.2 – 5.0 cP
Sample capacity
132 x 2 mL vial (two trays
default)
100 x 2 mL vial (two classic
trays optional)
36 x 6 mL vials (two trays
optional)