Agilent InfinityLab LC Series Vialsamplers User Manual
162
4
Preparing the Module
Setting up the Vialsampler
Auxiliary
•
Draw Speed
: determines the rate at which the plunger
draws sample from the vial. Set the speed to an appro-
priate value for your sample. For viscous samples, use
a slow Draw Speed.
•
Eject Speed
: determines the rate at which the plunger
ejects sample from the metering device. If you are
injecting large volumes of sample, setting a high Eject
Speed will shorten the time needed for an injection
cycle. For viscous samples in combination with multi-
draw, use a slow Eject Speed.
•
Wait Time After Draw
: ensures that the temporary vac-
uum, which originates from the drawing of liquid from
the sample vial, dissipates. The needle stays first on the
seat for the specified time, then after drawing sample
from the vial, remains there for the specified time.
•
Needle Height Offset
: a vertical offset that enables you
to position the needle a specific distance (in mm) away
from its standard position. The Offset function is useful
when analyzing very small sample volumes, or when
only a specific part of the sample is required, for exam-
ple, the top layer.
Table 17
Method parameter settings