Agilent InfinityLab LC Series Vialsamplers User Manual
121
3
Using the Module
Install the Optional Sample Cooler/Sample Thermostat
CAUTION
Damage to the cables
Do not bend or pinch the cables.
Make sure that the Sample Cooler/Sample Thermostat
fits perfectly in the sampler.
7
Slide the Sample Cooler/Sample Thermostat all the way
into the sampler.
8
Fix the Sample Cooler/Sample Thermostat with the four
screws.
9
Use a bubble level to check the leveling of the sampler.
NOTE
To ensure adequate drainage for condensate, the module
should be operated in a proper horizontal position.
10
Adhere a tubing holder clamp to the side of the
cooler/thermostat where the drain pipe is situated.
Ensure a distance of 60 mm from the bottom edge.
60 mm