Introduction
Overview of the Module
The Agilent InfinityLab Integrated Column Compartment, an optional upgrade for
the Vialsampler, eliminates the need for having a standalone column thermostat
in the LC stack. It has the heating capability of reaching temperatures up to 80 °C
and can hold a total of two columns with maximum lengths of 30 cm.
For applications requiring control over the vial temperature, the Vialsampler can
be equipped with the Agilent InfinityLab Sample Thermostat (G7167-60101). It
features a vapor-compression refrigeration system and an electric heater,
allowing the Vialsampler to reach vial temperatures down to 4 °C and up to 40 °C.