Transporting the Sampler
Prepare a Sampler with Cooler/Thermostat for
Transportation
When
The module needs to be transported or relocated.
WARNING
Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy
objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
CAUTION
Condensate inside the cooler or thermostat
Damage to the electronics
Unplug the power cords.
Drain off all condensate before dismounting the sample cooler or
thermostat.
Make sure that there is no condensate left.
CAUTION
Unsecured transportation
Mechanical damage
Secure the transport assembly before transporting the sampler.
NOTE
Moving the sampler with the Sample Cooler/Sample Thermostat installed is
possible for short distances (for example, from one workbench to another). For
longer transportation, remove the cooler/thermostat from the sampler and
handle the units separately.