Agilent InfinityLab LC Series Vialsamplers User Manual
208
7
Error Information
Vialsampler Error Messages
Cleanup after abort failed
Error ID: 34310
This, similarly to the
Command failed error
events, is a non-specific error message
and indicates only that something went wrong during the cleanup procedure. The
cleanup procedure refers to the process, which is executed when an analysis is
aborted (for example, due to a missing vial) and the components of the sampler
(such as the needle, gripper arm, and the metering device) need to be reset to the
initial conditions. As for the
Command failed error
events, this error message is
usually also coupled with other error events, which provide more specific
information on the nature of the issue. To identify these source error events, look
for entries with the same timestamp as for the
Cleanup after abort failed
error
message.
Wash procedure failed
Error ID: 34314
This, similarly to the
Command failed error
events, is a non-specific error message
and indicates only that something went wrong during the wash procedure. It is
usually coupled with other error events, which provide more specific information
on the nature of the issue. To identify these source error events, look for entries
with the same timestamp as for the
Wash procedure failed
error message.
Probable cause
Suggested actions
1
Cleanup procedure failed
Identify the source error message and follow the
instructions provided for that.
Probable cause
Suggested actions
1
Wash procedure failed
Identify the source error message and follow the
instructions provided for that.