Agilent InfinityLab LC Series Vialsamplers User Manual
177
6
Troubleshooting and Diagnostics
Status Indicators
Status Indicators
The Agilent InfinityLab Series LC modules are equipped with two status indicator
LED lights to enable the user to get an immediate visual impression of the actual
state of the instrument:
•
The power indicator light is integrated into the main switch of the module and
provides information about whether or not the system is powered on. When
the indicator illuminates in green, the module is ON.
•
The module status indicator light is situated in the upper right corner of the
module and provides information on the actual operating state of the system
(see
on page 58 for more information).