Agilent InfinityLab LC Series Vialsamplers User Manual
182
6
Troubleshooting and Diagnostics
Maintenance and Troubleshooting Tools
Using the Change Gripper Function
The
Change Gripper
function grants easy access to the release mechanism of the
gripper assembly by bringing the gripper arm into the front position.
The execution of this function encompasses the following elementary steps
when the Agilent Lab Advisor Software is used:
1
Select
Maintenance Positions
from the list under
Service & Diagnostics
.
2
Under
Task Selection
, select
Change Gripper
and click
Start
.
This gripper is moved into the maintenance position.
3
Replace the gripper arm (see
4
Click
Next
to move the gripper arm back to the home position.
5
Select
Back
in the bottom right corner to exit the
Maintenance Positions
function.