Agilent InfinityLab LC Series Vialsamplers User Manual
259
8
Maintenance
Exchange the Gripper Arm
3
Identify the slot for blocking the gripper arm rotation on
the left side of the transport arm (1) and the gripper arm
release button (2).
4
Place the hex key (1.5 mm) into the slot and rotate the
gripper arm slightly to the left or right until you find the
block position.
5
Press in the gripper arm release button and carefully
rotate the gripper arm approximately 40 – 50 ° to the
right, while firmly holding the hex key in place.
You should feel that the lock mechanism releases the
gripper arm.
6
Remove the gripper arm and the hex key.
NOTE
With firmware version D.07.28 or older, it is recommended
to turn off the sampler to complete the rest of the proce-
dure.
1
2
1.
2.