Agilent InfinityLab LC Series Vialsamplers User Manual
135
3
Using the Module
Transporting the Sampler
Install the Transport Protection Foam
When
To secure the transport arm before transporting or shipping the sampler.
Parts required
p/n
Description
G7129-40050
Transport Protection Foam
Preparations
All vials are removed from the vial drawers.
1
Open the doors of the module.
2
Move the gripper arm into the park position (see
on page 186) and turn off the sampler.
3
Place the Transport Protection Foam into the sampler.
4
Push the Transport Foam in until the endpoint.