Error Information
Sample Cooler/Sample Thermostat Error Messages
Sample temperature control switched off due to
overpressure
Error ID: 30716
The pressure in the refrigerant circuit exceeded the maximum allowed level. To
prevent any damage to the system, the compressor was turned off.
Sample temperature control sensor electronics
calibration failed
Error ID: 30717
The system is in the error state because the calibration of the analog
temperature sensor has failed.
Probable cause
Suggested actions
1
Overheated condenser
Turn off the cooler/thermostat and wait for
15 min to allow the system to cool down. Verify if
there is enough space around the sampler for
adequate ventilation and the cooler/thermostat
is not exposed to direct sunlight.
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Sampler incompatibility
If the hosting sampler is a Vialsampler, verify its
compatibility with the Sample Cooler installed.
Units with the serial number DEBAT02000 or
below are equipped with an analog temperature
sensor that is not compatible with the Vialsam-
pler.
2
Potential hardware error
Please contact your Agilent service representa-
tive.