Agilent InfinityLab LC Series Vialsamplers User Manual
134
3
Using the Module
Transporting the Sampler
1
Turn off the Sample Cooler/Sample Thermostat.
2
Remove condensate.
a
Place a suitable container underneath the outlet pipe.
b
Remove the drainage tube.
c
Gently tap the sides of the sampler several times to facilitate the drainage
of the condensate from the system.
3
Remove all vials from the vial drawers.
4
Move the transport assembly to the park position using Instant Pilot or Lab
Advisor, see
5
Turn off the sampler.
6
Install the Transport Protection Foam, see
“Install the Transport Protection
7
Remove the Sample Cooler/Sample Thermostat from the sampler if needed,
see
“Replace the Sample Cooler/Sample Thermostat”
NOTE
Do not tilt the module to avoid damage to the internal parts.