Agilent InfinityLab LC Series Vialsamplers User Manual
35
2
Site Requirements and Specifications
Site Requirements
Power Cords
Country-specific power cords are available for the module. The female end of all
power cords is identical. It plugs into the power-input socket at the rear. The male
end of each power cord is different and designed to match the wall socket of a
particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is
suitable for your particular country or region.
WARNING
Unintended use of power cords
Using power cords for unintended purposes can lead to personal injury or
damage of electronic equipment.
Never use a power cord other than the one that Agilent shipped with this
instrument.
Never use the power cords that Agilent Technologies supplies with this
instrument for any other equipment.
Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
WARNING
Absence of ground connection
The absence of ground connection can lead to electric shock or short
circuit.
Never operate your instrumentation from a power outlet that has no
ground connection.
WARNING
Electrical shock hazard
Solvents may damage electrical cables.
Prevent electrical cables from getting in contact with solvents.
Exchange electrical cables after contact with solvents.