Agilent InfinityLab LC Series Vialsamplers User Manual
256
8
Maintenance
Replace the Analytical Heads
3
Disconnect the two capillaries from the metering
device.
4
Remove the two fixing bolts.
5
Remove the metering head assembly from the sampler.
6
Install the metering head assembly in the autosampler.
Ensure that the drain hole of the metering body is facing
downwards. Tighten the bolts alternately two turns at a
time until the metering device is secure.
7
Reinstall the capillaries.
8
In Lab Advisor, select
Next
to move the piston back into
its normal position and then
Back
to finish the mainte-
nance procedure.
OR
In the Local Controller, select
Continue
to move the pis-
ton back into its normal position and then
Exit
to finish
the maintenance procedure.
9
Select
Back
to finish the procedure.
1.
2.
2.