Agilent InfinityLab LC Series Vialsamplers User Manual
242
8
Maintenance
Exchange the Sample Loop Assembly
1
In the Agilent Lab Advisor software select
Service &
Diagnostics> Maintenance Positions> Change Needle,
Loop and Seat
, click
Start
and wait until the needle
assembly is in maintenance position.
OR
In the Local Controller start the maintenance mode and
select
Change Needle, Loop and Seat
function.
2
Open the doors of the module.
3
Loosen the safety cover screw.
WARNING
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
4
Remove the safety cover.