Maintenance
Exchange the Rotor Seal
7
Install the stator ring with the short of the two pins fac-
ing towards you at the 12 o'clock position. Ensure the
ring sits flat on the valve body.
8
Install the stator head. Tighten the bolts alternately two
turns at a time until the stator head is secure.
NOTE
Do not over-tighten the screws. The screws hold the assem-
bly together and do not affect the sealing force. The sealing
force is automatically set as the screws close the stator
head against the valve body.