Configuring the Remote Interface
Programming Reference
E1852B Operating Guide
105
Configuring the Remote Interface
Interface selection
You can choose to control the test set remotely using either the GPIB or
parallel interfaces.
Use the rear panel DIP switch to select the remote interface type. Refer
to Rear Panel Connections on page 13 or the E1852B Bluetooth Test
Set
Installation Guide
for more detail.
GPIB Address
Each device on the GPIB (IEEE-488) interface must have a unique
address. The address is set on the rear panel DIP switches A1 to A5.
The switches are used to set the address in binary format. A1 is the
least significant bit, A5 the most. You can set the test set address to any
value between 0 and 31.
Your GPIB bus controller has its own address. Avoid using the bus
controller’s address for any instrument on the interface bus.
Controllers generally use address 21.
A specific address is not required for parallel port operation.
Factory Default
When shipped from the factory, the DIP switches are configured for
address 15.