Detailed Command Descriptions
Programming Reference
E1852B Operating Guide
119
CONFigure:RF:DIRTY_TX:OFFSet <value>
Control
CONFigure:RF:EXT_ATT <numeric value>
External RF attenuation compensation
Syntax:
CONFigure:RF:DIRTY_TX:OFFSet <value>
Value Range:
-75 to 75
[kHZ]
Default Value:
0
Applicable in
State:
Set: All
Query: All
Description:
Use this command to specify the dirty transmitter frequency offset
impairment. The offset is disabled when the
RF:DIRTY_TX
<value>
is set to
OFF
.
Examples:
CONF:RF:ATT:DIRTY_TX:OFFS 50
Sets a 50 kHz
offset
CONF:RF:ATT:DIRTY_TX:OFFS?
Returns current
setting
Syntax:
CONFigure:RF:EXT_ATT <numeric value>
Value Range:
1 to 100
[dB]
Default Value:
0.0
Applicable in
State:
Set: All
Query: All
Description:
This command states the amount of attenuation, caused by cable
and connectors, applied to the RF signal between the test set and
the DUT. The test set compensates for this attenuation in the
measurement results.
Examples:
CONF:RF:EXT_ATT 0.2
Sets 0.2 dB
CONF:RF:EXT_ATT?
Returns current
setting
E1852-90007.book Page 119 Monday, February 4, 2002 2:09 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com