Detailed Command Descriptions
Programming Reference
E1852B Operating Guide
125
CONFigure:RFGEnerator:TUNE <value>
RF Generator Tuning Frequency
CONFigure:SAMPLE:MODE <setting>
Slave Bluetooth Device Address (DUT)
Syntax:
CONF:RFGEnerator:TUNE <value>
Value Range:
-3
Offset =
−
300KHz
Default Value:
-2
Offset =
−
200KHz
0
-1
Offset =
−
100KHz
0
Offset = 0KHz
1
Offset = 100KHz
2
Offset = 200KHz
3
Offset = 300KHz
Applicable in
State:
Set: All
Query: Not Applicable
Description:
This command enables fine tuning of the RF Generator frequency,
with the specified offset. Only used for special purposes.
Example:
CONF:RFGE:TUNE
−
2
Sets a
−
200 kHz offset
Syntax:
CONFigure:SAMPLE:MODE <setting>
Value Range:
AUTO
Sampling handled
automatically
Default Value:
CTRL
AUTO
Applicable in
State:
Set: All
Query: All
Description:
This command configures the sampling mode of the test set. When
AUTO
is selected sampling is occurs when a new measurement is
requested. When
CTRL
is selected, sampling only occurs when
triggered by the
PROC:SAMPLE:FORCE
command.
Examples:
CONF:SAMPLE:MODE CTRL
Sets triggered
mode
CONF:SAMPLE:MODE?
Returns current
setting
E1852-90007.book Page 125 Monday, February 4, 2002 2:09 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com