Programming Reference
Detailed Command Descriptions
120
E1852B Operating Guide
CONFigure:RF:LEVel <numeric value>
RF output level of the test set
Syntax:
CONFigure:RF:LEVel <numeric value>
Value Range:
-85 to 0
sets test set RF output [dBm] Default Value:
−
35 dBm
Applicable in
State:
Set: All
Query: All
Description:
This command configures the output power of the test set. When
used in the RF Generator mode, (RFGE) it is necessary to wait
approximately 200ms. before the new output level is valid.
Examples:
CONF:RF:LEV –60
Sets RF level
to
−
60 dBm
CONF:RF:LEV?
Returns current
setting
E1852-90007.book Page 120 Monday, February 4, 2002 2:09 PM