Programming Reference
Detailed Command Descriptions
132
E1852B Operating Guide
PROCedure:INQuiry:STOP
Halt the current inquiry scan
PROCedure:NACK:START <numeric value>
Start NACK Count (Packet Error Rate)
N O T E
This command erases all previous NACK data and starts a new
measurement. This command should hence be used when different
parameters are configured during a NACK test.
Syntax:
PROCedure:INQuiry:STOP
Value Range:
Not Applicable
Default Value:
Not Applicable
Applicable in
State:
Set: IDLE, BT, TESTMODE
Query: Not Applicable
Description:
This command halts the current inquiry to find any Bluetooth
devices in the vicinity of the test set.
Example:
PROC:INQ:STOP
Halts the inquiry
scan
Syntax:
PROCedure:NACK:START <numeric value>
Value Range:
1 to 1.6E6
Number of packets used for
calculating the NACK count
Default Value:
Not Applicable
Applicable in
State:
Set: CONN, TCONN
Query: Not Applicable
Description:
This command is used to start an unacknowledged (NACK) count
measurement. A NACK measurement can be used as an
alternative to the BER measurement. The NACK measurement
can be made in Normal mode, whereas you can only make a BER
measurement in Testmode.
All unacknowledged packets are counted and compared the total
number of packets sent.
Example:
PROC:NACK:START 1000
Sets a value of
1000 packets
E1852-90007.book Page 132 Monday, February 4, 2002 2:09 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com