READ:DELTA_F2_F1_RATIO?
Query Delta F2 average versus Delta F1 average ratio
Syntax:
READ:DELTA_F2_F1_RATIO?
Value Range:
<value>
Default Value:
Not Applicable
Applicable in
State:
Query: TCONN, RFGE
Description:
This query is used to retrieve the current ratio between the Delta
F2 average and the Delta F1 average, as specified in the Bluetooth
Test Specification version 0.9, chapter 5.1.9. The most recent
values of Delta F1 and Delta F2 averages are used in the
calculation (the last results from the queries
READ:DELTA_F2?
and
READ:DELTA F1?
For a valid result both BS55 and BS0F modulation must be applied
in test mode
Example:
READ:DELTA_F2_F1 RATIO?