PC User Interface
System Page
26
E1852B Operating Guide
System Page structure
The user interface always opens with the
System
page displayed. It
contains the following panels:
1. Communication Port
The radio buttons on the
Communication Port
panel are used to select
the PC port you want to use for control of the test set.
2. GPIB Configuration
The
GPIB-Board Manufacturer
scroll button is used to identify the
type of card you have fitted to your computer. If you have an Agilent
card, select
7
using the
Board ID
scroll buttons, otherwise use
0.
Select the GPIB address of the test set (default factory setting 15) using
the
Address
scroll buttons.
3. Firmware Update (Debug Mode Only)
The Firmware Update panel is used to set the required communication
port and initiate the firmware download process. This is only available
in debug mode and you should refer to the instructions on the E1852B
Software Home Page. (The URL can be found on the
About
page of the
user interface.)
4. Wait in Testmode Setup
This command sets the test set to wait in test mode before sending the
test activate message to the DUT. This is required by some DUTs. It
does not apply to a normal mode connection