Programming Reference
Detailed Command Descriptions
112
E1852B Operating Guide
CONFigure:AUDIO_MUTE:TX <value>
Control audio in the transmit direction
CONFigure:AVERage:BURst <value>
Number of bursts used for Measurements
Syntax:
CONFigure:AUDIO_MUTE:TX <value>
Value Range:
ON
Mutes audio in the receive
direction
Default Value:
OFF
OFF
Un-mute audio in the receive
direction
Applicable in
State:
Set: All
Query: All
Description:
This command is used to mute or un-mute the audio signal in the
transmit direction.
Examples:
CONF:AUDIO_MUTE:TX OFF
Sets mute off
CONF:AUDIO_MUTE:TX?
Returns current
setting
Syntax:
CONFigure:AVERage:BURst <value>
Value Range:
1 to 200
Number of bursts
Default Value:
1
Applicable in
State:
Set: All
Query: All
Description:
This command is used to configure the number of bursts used for
averaging when measuring NTP, PTP, Freq.Offset, Freq.Drift and
Deviation.
A higher number gives more stable and precise measurements but
also increases the duration of the measurements.
Examples:
CONF:AVER:BUR 10
Sets loopback
test mode
CONF:AVER:BUR?
Returns current
setting
E1852-90007.book Page 112 Monday, February 4, 2002 2:09 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com