Programming Reference
Detailed Command Descriptions
154
E1852B Operating Guide
SYSTem:TEST:ADC_BUF?
Self test of ADC Buffer
SYSTem:TEST:COMM_BUF?
Self test of Communication Buffer
Syntax:
SYSTem:TEST:ADC_BUF?
Value Range:
OK
Error, xx
Test is OK
Test failed. xx is the number
of errors detected
Default Value:
Not Applicable
Applicable in
State:
Query: IDLE
Description:
This selftest command tests that the PC is able to access the ADC
buffer of the test set.
Example:
SYST:TEST:ADC_BUF?
Syntax:
SYSTem:TEST:COMM_BUF?
Value Range:
OK
Error, xx
Test is OK
Test failed. xx is the number
of errors detected
Default Value:
Not Applicable
Applicable in
State:
Query: IDLE
Description:
This selftest command tests that the PC is able to access the
communication buffer of the test set.
Example:
SYST:TEST:COMM_BUF?