PC User Interface
Normal Mode Page
50
E1852B Operating Guide
Normal Mode Page structure
The
Normal Mode
page contains the following panels and entry fields:
1. RF Level
The RF power level in the test set transmit direction can be adjusted
for sensitivity measurements.
The power levels can be adjusted from –85 to 0dBm.
The power level can be entered directly in the
RF Level [dBm]
field or
adjusted by moving the
RF Level
slider.
2. Meas. Averaging Factor
You can choose to make measurements on single packets or average the
results over a number of packets up to a maximum of 999. A lower
number updates the measurement more frequently.
3. Audio
The test set currently supports the CVSD audio CODEC.
4. DUT Power Control
If the DUT supports this feature, you can use the up/down buttons to
control its RF output level.
5. Packet Error Rate (PER)
The running Packet Error Rate is displayed on the Normal Mode
Measurements
window.