98
Chapter 7
Processing Measurement Results
Reading Measurement Data at Specific Measurement Points (Using the
Marker Function)
Reading the marker-specified value
Use the following commands to read the marker-specified measurement or sweep
parameter value.
You can use the following command to display a list of all marker-specified values on the
Agilent 4294A LCD. Note, however, that you cannot use these commands to read these
values.
•
“MKRL” on page 352
Sample program for using the marker function
Example 7-1 shows a sample program for using the marker function. This program is
available under the filename of “marker.bas” on the sample program disk.
This program performs sweep once, uses the marker function to search for the maximum
peak and the minimum negative peak in the 10-100 MHz range, provides a submarker at
each of these peaks, and reads and displays the submarker-specified values.
Lines 30 to 70
Identifies the external controller and Instrument BASIC and sets the
GPIB address.
Lines 90 to 110
Substitutes the lower and upper border values for the search range and
the trace selected as the active trace into the corresponding variables
Sear_min$, Sear_max$, and Act_trc$, respectively.
Lines 130 to 160 Resets the Agilent 4294A, performs a sweep once, and waits for the
sweep to be completed.
Lines 180 to 190 Sets the trace selected as the active trace to Act_trc$ and executes the
auto scale to automatically set the scale parameters so that the
waveforms fit on the screen.
Lines 210 to 240 Turns ON the marker and sets the lower and upper border values for
the search range to Sear_min$ and Sear_max$, respectively.
Lines 260 to 300 Defines the peak by specifying “0.1/1e6” and “positive” as peak slope
threshold and peak polarity, searches for the peak and moves the
marker to that peak, and positions submarker 1 at the peak.
Lines 320 to 340 Redefines the peak by specifying “negative” as peak polarity, searches
Measurement Parameter
Value
Supplementary Reading for
Measurement Parameter
Value (Vector)
Sweep Parameter Value
Marker
“OUTPMKR?” on page 387
“MKRVAL?” on page 360
“MKRAUV?” on page 351
“MKRPRM” on page 358
Used for query
Submarker “OUTPSMKR{1-7}?” on page 393
“SMKRVAL{1-7}?” on
page 441
“SMKRAUV{1-7}?” on page
438
“SMKRPRM{1-7}” on page 440
Used for query
Δ
Marker
“OUTPDMKR?” on page 377
“DMKRVAL” on page 306
Used for query
*1
“DMKRAUV” on page 303
Used for query
*1
“DMKRPRM” on page 305
Used for query
*1.You can read parameter values only when you use the fixed
Δ
marker.
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...