Chapter 16
357
GPIB Command Reference
MKRP
16. GPI
B
Com
m
an
d Re
fere
nce
( F -
N )
MKRP
Syntax
MKRP <numeric>
MKRP?
Description
For the active trace (set with the “TRAC” command on page 457), moves the marker to the
specified measurement point. In the coupled marker mode (set with the “MKRCOUP”
command on page 352), regardless of the active trace setting, moves the markers on both
the A and B traces. If this command is executed as Query, it reads out the measurement
point number at the marker position. If the marker function is off, executing this command
causes an error and the command is ignored.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
In the continuous marker mode, the number of the largest measurement point that does not
exceed the marker position is reads out.
Corresponding
key
No front panel key is available to execute this function.
MKRPKD
Syntax
MKRPKD
Description
For the active trace (set with the “TRAC” command on page 457), sets
Δ
X and
Δ
Y of the
peak definition so that the current marker position is recognized as a peak. In other words,
Δ
X and
Δ
Y are respectively set sweep parameter value difference and measurement
parameter value difference between the maker and its left adjacent measurement point.
Note that, if the
Δ
maker mode is on,
Δ
X and
Δ
Y are respectively set sweep parameter
value difference and measurement parameter value difference between the maker and the
Δ
maker.
If the marker function is off, executing this command causes an error and the command is
ignored. (No query)
Corresponding
key
[Search]
-
PEAK
-
PEAK DEF MENU
-
MKR
→
PEAK DELTA
<numeric>
Description
Number of the measurement point to which the marker is moved
(the number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
Range
1 to the number of points
Resolution
1
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...