Chapter 16
301
GPIB Command Reference
DISP
16. GPI
B
Com
m
an
d Re
fe
re
nce
( * , A - E
)
DISP
Syntax
DISP {DATA|MEMO|DATM|DMNM|DELP|DDVM}
DISP?
Description
For the active trace (set with the “TRAC” command on page 457), makes the setting
related to the displayed trace.
If a parameter other than DATM is set, the scale setting target trace (set with the “SCAF”
command on page 422) and the marker use target trace (set with the “MKRO” command
on page 356) are also changed.
If no data has been copied to the memory trace (the “DATMEM” command on page 285
has not been executed), executing this command to set a parameter other than DATA
causes an error and the command is ignored.
If the measurement parameter is a scalar value, executing this command to set DDVM
causes an error and the command is ignored. If the measurement parameter is a vector
value, executing this command to set DELP causes an error and the command is ignored.
Parameters
Query response
{DATA|MEMO|DATM|DMNM|DELP|DDVM}<newline><^END>
Corresponding
key
[Display]
-
DEF TRACE[ ]
- {
DATA
|
MEMORY
|
DATA and MEMORY
|
DATA-MEM
|
DELTA %
|
DATA/MEM
}
Description
*1
DATA
(initial value)
Displays the data trace. The contents of the data trace is measured
data (no operation).
MEMO
*2
Displays the memory trace.
DATM
*2
Displays both the data trace and the memory trace at the same time.
The contents of the data trace is measured data (no operation).
DMNM
*2
Displays the data trace. The contents of the data trace is the
operation result of DATA - MEM.
DELP
*2,*3
Displays the data trace. The contents of the data trace is the
operation result of (DATA - MEM) / MEM.
DDVM
*2,*4
Displays the data trace. The contents of the data trace is the
operation result of DATA / MEM.
*1.In operation expressions in the description, DATA indicates measured
data and MEM indicates data of the memory trace.
*2.Selectable only when data has been copied in the memory trace.
*3.Selectable only when the measurement parameter is a scalar value.
*4.Selectable only when the measurement parameter is a vector value.
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...