Chapter 16
321
GPIB Command Reference
INPUCOMC{1-3}
16. GPI
B
Com
m
an
d Re
fere
nce
( F -
N )
INPUCOMC{1-3}
Syntax
INPUCOMC{1-3} <numeric 1>,<numeric 2>,..,<numeric NOP
×
2-1>,<numeric NOP
×
2>
Description
Enters the OPEN/SHORT/LOAD compensation data for fixture compensation, calculates
the compensation coefficient, and stores it into the non-volatile memory (backup memory).
The “INPUCOMC1” command is for OPEN compensation data (G-B); the
“INPUCOMC2” command, SHORT compensation data (R-X); the “INPUCOMC3”
command, for LOAD compensation data (R-X).
Enter data after setting the standard values, because the standard values to be used in
measuring data is needed to calculate fixture compensation coefficients.
A mismatch, between the number of required data at the execution of the command
(defined by the measurement point setting for fixture compensation data (set with the
“CALP” command on page 275) and the number-of-points setting) and the number of
entered parameters, causes an error and the command is ignored. (No query)
Parameters
Total of NOP
×
2 parameters are required because the fixture compensation data value
(complex number) of each measurement point is divided into 2 parameters for its real part
and imaginary part. Where, NOP is the number of points (for user measurement points) or
100 (for fixed measurement points), and n is an integer between 1 and NOP.
Corresponding
key
No front panel key is available to execute this function.
INPUDATA
Syntax
INPUDATA <numeric 1>,<numeric 2>,..,<numeric NOP
×
2-1>,<numeric NOP
×
2>
Description
Enters a data array (refer to “Internal data arrays” on page 81). A mismatch, between the
number-of-points setting at the execution of the command and the number of entered
parameters, causes an error and the command is ignored. (No query)
Parameters
Total of NOP
×
2 parameters are required because the measurement parameter value of each
measurement point (complex number) is divided into 2 parameters for its real part and
imaginary part. Where, NOP is the number of points, and n is an integer between 1 and
NOP.
Corresponding
key
No front panel key is available to execute this function.
Description
<numeric n
×
2-1>
Real part of the compensation data value (complex number) of the
n-th measurement point.
<numeric n
×
2>
Imaginary part of the compensation data value (complex number) of
the n-th measurement point.
Description
<numeric n
×
2-1>
Real part of the measurement parameter value (complex number) of
the n-th measurement point.
<numeric n
×
2>
Imaginary part of the measurement parameter value (complex
number) of the n-th measurement point.
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...