![Agilent Technologies 4294A Скачать руководство пользователя страница 462](http://html.mh-extra.com/html/agilent-technologies/4294a/4294a_programming-manual_2867882462.webp)
462
Chapter 16
GPIB Command Reference
WIDVTYPE
WIDVTYPE
Syntax
WIDVTYPE {DIVS2|MULS2|DIV2|FIXed}
WIDVTYPE?
Description
For the active trace (set with the “TRAC” command on page 457), selects the method to set
a cutoff point in the trace bandwidth analysis function.
Parameters
If the measurement parameter is a vector value (for COMPLEX Z-Y), you can select
FIXed only.
Query response
{DIVS2|MULS2|DIV2|FIXed}<newline><^END>
Corresponding
key
[Search]
-
WIDTHS DEF MENU
- {
MKRVAL/(
√
2)
|
MKRVAL*(
√
2)
|
MKRVAL/2
|
FIXED VALUE
}
WOPEN
Syntax
ROPEN <string>
Description
If the specified file exists, this command makes it write-enabled; otherwise, creates a new
file and makes it write-enabled. This command takes its arguments in a different way,
depending on the file format.
The size of an existing file cannot be changed. Therefore, if you want to change them,
delete the file itself using the “PURG”(406page) command and then create a new file using
this command.
This command is used in combination with the “WRITE”(463page) command and the
“CLOSE”(280page) commands, as shown in Figure 16-8 on page 412. (No query)
Parameters
Corresponding
key
No front panel key is available to execute this function.
Description
DIVS2
The value obtained by dividing the measurement parameter value
at the marker position by
is set as the cutoff point.
MULS2
The value obtained by multiplying the measurement parameter
value at the marker position and
is set as the cutoff point.
DIV2
The value obtained by dividing the measurement parameter value
at the marker position by 2 is set as the cutoff point.
FIXed (initial
value)
Lets you set the cutoff point freely using the “WIDFVAL”
command on page 460.
2
2
Description
<string>
File name of up to 12 characters including its extention
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...