Chapter 16
393
GPIB Command Reference
OUTPSMKR{1-7}?
16. GPI
B
Com
m
an
d Re
fe
re
nce
( O -
R )
Figure 16-6
Analyzed parameters
OUTPSMKR{1-7}?
Syntax
OUTPSMKR{1|2|3|4|5|6|7}?
Description
For the active trace (set with the “TRAC” command on page 457), reads out the
measurement parameter value and sweep parameter value at the sub marker (marker
number 1 to 7) position. If the sub marker is off, executing this command causes an error
and the readouts are invalid. (Query only)
Query response
{numeric 1},{numeric 2},{numeric 3}<newline><^END>
Readouts are as follows:
{numeric 1}:
Readout of the measurement parameter value at the sub marker
position. If the measurement parameter is a scalar value (for other than
COMPLEX Z-Y), the measurement parameter value is read out. If the
measurement parameter is a vector value (for COMPLEX Z-Y), the
readout is as follows.
•
If the polar chart format is selected, the amplitude is read out.
•
If the complex plane format is selected, resistance (for trace A) or
conductance (for trace B) is read out.
{numeric 2}:
Subsidiary readout of the measurement parameter value at the sub
marker position. If the measurement parameter is a scalar value (for
other than COMPLEX Z-Y), 0 is always read out. If the measurement
parameter is a vector value (for COMPLEX Z-Y), the readout is as
follows:
•
If the polar chart format is selected, the phase is read out.
•
If the complex plane format is selected, reactance (for trace A) or
susceptance (for trace B) is read out.
{numeric 3}:
Sweep parameter value at the sub marker position.
Corresponding
key
No front panel key is available to execute this function.
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...