Chapter 16
267
GPIB Command Reference
ANARANG
16. GPI
B
Com
m
an
d Re
fe
re
nce
( * , A - E
)
ANARANG
Syntax
ANARANG <numeric 1>,<numeric 2>[HZ|MHZ|V|A]
ANARANG?
Description
Sets the analysis range when performing waveform analysis using the waveform analysis
command. You specify the analysis range using the lower limit and the upper limit (or, the
upper limit and the lower limit). If the lower/upper limit does not match with any
measurement point sweep parameter values, the analysis range lower limit is set to the
minimum measurement point larger than the specified value and the analysis range upper
limit is set to the maximum measurement point smaller than the specified value.
If the sweep condition setting is changed, the analysis range is set to the entire sweep
range. If the sweep range setting is zero span, the analysis range is always the entire sweep
range (all measurement points).
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric 1},{numeric 2}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric 1>
<numeric 2>
Description
Analysis range lower
limit (or upper limit)
Analysis range upper
limit (or lower limit)
Range
Current sweep range
Current sweep range
When the sweep
parameter is
frequency
Initial value 40
110E6
Unit
Hz
Hz
Resolution
1E-3
1E-3
When the sweep
parameter is OSC
level (voltage)
Initial value 0
1
Unit
V (volt)
V (volt)
Resolution
1E-3
1E-3
When the sweep
parameter is OSC
level (current)
Initial value 40E-6
20E-3
Unit
A (ampere)
A (ampere)
Resolution
1E-6
1E-6
When the sweep
parameter is dc bias
level (voltage)
Initial value 0
0
Unit
V (volt)
V (volt)
Resolution
1E-3
1E-3
When the sweep
parameter is dc bias
level (current)
Initial value 0
0
Unit
A (ampere)
A (ampere)
Resolution
20E-6
20E-6
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...