Chapter 16
355
GPIB Command Reference
MKRMON
16. GPI
B
Com
m
an
d Re
fere
nce
( F -
N )
MKRMON
Syntax
MKRMON {OFF|ACV|ACI|DCV|DCI}
MKRMON?
Description
Enables/disables the display of the oscillator (OSC) level or dc bias level monitor value at
the marker position measurement point. If you try to specify monitoring an item whose
level monitor has been disabled in the oscillator level monitor function (set with the
“OMON” command on page 365) or dc bias level monitor function (set with the “BMON”
command on page 272), executing this command causes an error and the command is
ignored.
Parameters
Query response
{OFF|ACV|ACI|DCV|DCI}<newline><^END>
Corresponding
key
[Utility]
-
LEVEL MON [ ]
- {
OFF
|
AC-V
|
AC-I
|
DC-V
|
DC-I
}
MKRMOV
Syntax
MKRMOV {CONT|DISC}
MKRMOV?
Description
For the active trace (set with the “TRAC” command on page 457), switches between the
continuous marker mode (markers can be moved to between measurement points) and the
discrete marker mode (markers can be to measurement points only). In the coupled marker
mode (set with the “MKRCOUP” command on page 352), regardless of the active trace
setting, this setting is applied to both the A and B traces. In the continuous marker mode,
the marker readout is an interpolated value derived from the data of the measurement
points on both sides.
Parameters
Query response
{CONT|DISC}<newline><^END>
Corresponding
key
[Marker]
-
MKR [ ]
Description
OFF (initial value)
Turns off the marker level monitor function.
ACV
Specifies the monitor value display of the oscillator voltage level.
ACI
Specifies the monitor value display of the oscillator current level.
DCV
Specifies the monitor value display of the dc bias voltage level.
DCI
Specifies the monitor value display of the dc bias current level.
Description
CONT (initial value)
Specifies the continuous marker mode.
DISC
Specifies the discrete marker mode.
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...