54
Chapter 4
Preparing For Accurate Measurement
Adapter Setting
Adapter Setting
To select an adapter, use the command given below.
•
“E4TP” on page 309
If you select any adapter other than NONE, you need to measure the data indicated in the
table below to calculate the setup data.
To measure data, use the command given below.
•
“ECAL{P|A|B|C}” on page 309
After measuring the necessary data, execute the command given below to calculate the
setup data for the selected adapter and then store this in the nonvolatile memory
(EEPROM).
•
“ECALDON” on page 309
NOTE
If you execute the “ECAL{P|A|B|C}” command to measure unnecessary data, an error
occurs and the command is ignored.
To suspend the above process, execute the command given below.
•
“ECALQUI” on page 310
A sample program for adapter setting is shown in Example 4-1. This program is given the
file name adapter.bas and is stored on the sample program disk.
This program selects 7mm 42942A for adapter setting, measures phase data, open data,
short data and load data and then stores them in nonvolatile memory.
Lines 40 to 100
These lines distinguish between the external controller and Instrument
BASIC and set the GPIB address and the select code.
Line 110
This line substitutes the adapter selection to the variable Adapter$.
Lines 130 to 160 These lines reset the Agilent 4294A and then set the adapter selection
to the variable Adapter$.
Lines 180 to 190 These lines set 1 to Bit 8 of the instrument event status register (the bit
for completion of the measurement of calibration data and
compensation data) and to Bit 2 of the service request enable register
so that SRQ can be used.
Adapter
Phase
Open
Short
Load
NONE
Unnecessary
Unnecessary
Unnecessary
Unnecessary
4TP 1M
Necessary
Unnecessary
Unnecessary
Necessary
4TP 2M
Necessary
Unnecessary
Unnecessary
Necessary
7mm 42942A
Necessary
Necessary
Necessary
Necessary
PROBE 42941A
Necessary
Necessary
Necessary
Necessary
Содержание 4294A
Страница 4: ...4 ...
Страница 18: ...18 Contents ...
Страница 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Страница 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Страница 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Страница 130: ...130 Chapter9 Communication with External Equipment Using the I O Ports Sample Program to Use the I O Port ...
Страница 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Страница 246: ...246 13 Application Sample Programs File Transfer Function ...
Страница 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Страница 507: ...Index Index 507 Index ...
Страница 508: ...508 Index Index ...
Страница 509: ...Index Index 509 Index ...
Страница 510: ...510 Index Index ...
Страница 511: ...Index Index 511 Index ...
Страница 512: ...512 Index Index ...
Страница 513: ...Index Index 513 Index ...
Страница 514: ...514 Index Index ...
Страница 515: ...Index Index 515 Index ...
Страница 516: ...516 Index Index ...
Страница 517: ...Index Index 517 Index ...
Страница 518: ...518 Index Index ...