Monitor Configuration
The Precision Xceed Pro Monitor configuration settings are enabled via the data management system. These settings
can be viewed on the hand-held device, but cannot be changed on it. Settings that are configured through the data
management system are:
Upload interval
Length of time between data uploads.
•
Whether you can continue to test (“allow test” or “warn”) or not (“lockout”).
•
Operator ID
Whether the Operator ID is required, optional or not used for all tests.
•
Define the Operator ID prompt.
•
The minimum and maximum number (between 1 and 30) of digits that are acceptable for the Operator ID.
•
With manual entry of the Operator ID, determine whether a check digit is required (modulus 10 or 11).
•
Decide acceptable formats for the Operator ID bar code. Refer to the Bar Code Specifications in Chapter 12 for
•
available formats.
The monitor can truncate (ignore) the leading digits and/or the last digits of an Operator ID. The monitor can be set
•
to select specific digits of an Operator ID. The number of digits is configured in the data management system.
Patient Tests / Patient ID
Determine whether the Patient ID is required, optional or not used for patient tests.
•
Define the Patient ID prompt.
•
The minimum and maximum number (between 1 and 30) of digits that are acceptable for the Patient ID.
•
Permission for manual entry of the Patient ID.
•
Determine whether duplicate entry of a manually entered ID is required.
•
With manual entry of the Patient ID, determine whether a check digit is required (modulus 10 or 11).
•
Decide acceptable formats for the Patient ID bar code. Refer to the Bar Code Specifications in Chapter 12 for
•
available formats.
The monitor can truncate (ignore) the leading digits and/or the last digits of a Patient ID. The monitor can be set to
•
select specific digits of a Patient ID. The number of digits is configured in the data management system.
Decide whether confirmation of the Patient ID required.
•
Determine how the ID will be confirmed.
•
Decide whether comment code is required, optional or not used for patient tests that are in the acceptable range
•
for the facility and the test strip.
Decide whether comment code is required, optional or not used for patient tests that are outside of the acceptable
•
range for the facility and the test strip.
Control Tests
Set which levels of control solution are required, optional or not used by your facility. The monitor can be
•
configured to run any combination of low, mid and high control solution tests.
Set whether control solution test results display as numeric or as Pass/Fail.
•
Determine whether a comment code is required, optional or not used for control tests that are within the
•
acceptable range for the facility and the test strip.
Decide whether a comment code required, optional or not used for control tests that are outside of the acceptable
•
range for the facility and the test strip.
QC Interval
Determine how frequently QC tests need to be run. This can be set to a number of hours or up to three specified
•
times during the day.
Decide whether the operator can continue to test (“allow test” or “warn”) or not (“lockout”) when the QC interval
•
is expired.
Review Setup
Chapter 8: Review Setup
Содержание Precision Xceed Pro
Страница 37: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test β Ketone ...
Страница 42: ...7 Linearity Test ...
Страница 60: ...9 Data Upload Docking Station Mounting Instructions Enabling Data Upload ...
Страница 77: ...11 Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries ...
Страница 80: ...12 Specifications Precision Xceed Pro Monitor Symbols Bar Code Types Important Safety Instructions ...