What You See on the Display
What You Do
Comments
4-4
Questions? Refer to Page 10-10 for Technical Support Instructions.
9.
Wait for the monitor to analyze the
sample and display the test result.
The monitor counts down then displays
the test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
10.
Note the test result and whether it
falls within the acceptable range.
If required, scan or manually enter the
comment code, and press
Enter.
If there is no prompt to enter a
comment code, skip down to step 11.
Control test results can appear as numeric
or PASS/FAIL, depending on how the
test is set up. If numeric, an “X” will appear
in front of a test result that is out-of-range.
The monitor may be enabled to scan or enter
a 1- to 2-digit comment code.
11.
You can select one of the following:
Press
1 – Next Level
.
Press
2 – Repeat Test
.
Press
Menu
to return to the
Menu Mode
menu.
Press
On/Off
to turn off the monitor.
Remove the test strip from the monitor
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
The display shows:
Level
(Low, Mid, High)
Comment Code
(Number in box)
Test Type
(Glucose)
Date and Time of test
Operator ID
(
o:
prompt)
Note that test date and time alternates with
Operator ID.
If
1
is selected, the
Scan or Enter Control
Lot
screen will appear. Return to step 4 to
perform the
Next Level
control test.
If
2
is selected, the
Scan or Enter Strip
Lot
screen will appear. Return to step 5 to
Repeat Test
at the same level.
Note:
If β
-
ketone testing is disabled, then following a complete set of Glucose controls,
1-Next Level becomes 1-Exit.
If β
-
ketone testing is enabled and Glucose QC is run first, then following a complete set of
Glucose controls, 1-Next Level becomes 1-QC Ketone.
If β
-
ketone testing is enabled and β
-
ketone QC is run first, then following a complete set of
β
-
ketone controls, 1-Next Level becomes 1-QC Glucose.
99 mg/dL
Chapter 4 : Control Test
Содержание Precision Xceed Pro
Страница 37: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test β Ketone ...
Страница 42: ...7 Linearity Test ...
Страница 60: ...9 Data Upload Docking Station Mounting Instructions Enabling Data Upload ...
Страница 77: ...11 Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries ...
Страница 80: ...12 Specifications Precision Xceed Pro Monitor Symbols Bar Code Types Important Safety Instructions ...